Seleccione su idioma
Buenos Aires, Argentina: Perú 275, C1067AAE
Tel: (011) 4331.0707 Fax: (011) 4331.5917
E-mail: luftman@phelectronica.com.ar


Conducell 4UxF - Arc
Modelo: Conducell 4UxF - Arc
imagen producto

Conducell 4UxF Arc 4-pole conductivity sensors have a VP8 connector and are available with PG13.5 process connections. Arc technology enables the storage of sensor calibrations, predictive diagnostics, automated documentation, as well as user and process assignment. Direct connection to the control system eliminates the need for a transmitter and the many available communication protocols provide the flexibility for simple integration. FDA certified Conducell 4UxF sensors are CIP, SIP and autoclave compatible for easy use in hygienic applications with good linearity over a wide conductivity range. Electrodes are available in stainless steel (USF, 1.4435 and UHF, 2.4602), platinum (UPtF) and titanium (UTF). 1.4435 stainless steel is the most common electrode material and is suitable for typical biopharmaceutical applications. Platinum can be used to achieve long-term stability and resistance to corrosion. In highly corrosive or chemically aggressive applications, 2.4602 stainless steel can be chosen for its exceptional strength. Titanium is suitable for use with corrosion such as seawater.
Product Specifications
a-length 120 mm
Electrical Connector VP8
Measurement Principle 4 pole contacting
Measuring Range 1 µS/cm to 300 mS/cm
Accuracy at 25 °C ± 3 % at 1 µS/cm to 100 mS/cm, ± 5 % at 100 to 300 mS/cm
Cell Constant 0.36 /cm
Temperature Sensor NTC 22 kOhm
Configurable Values Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K Diameter 12 mm
Process Connection PG13,5
Wetted Parts Stainless Steel 1.4435 PEEK (Polyetheretherketone) EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document
Surface Quality of Steel Ra < 0.4 µm (N5, electropolished)
Analog Interface 1 4 to 20 mA for conductivity, programmable
Analog Interface 2 4 to 20 mA for Temp., programmable
Analog Interface 1 and 2 Galvanically not isolated; pulse width modulation 3.5 kHz
Digital Interface RS485 Modbus (max. 31 addresses)
Baud Rate 4800, 9600, 19200, 38400, 57600,115200 bd
Operating Voltage 7 to 30 VDC max. 150 mW
Serial Number Yes
Certificate Yes, with calibrated cell constant ATEX Approval No IECEx Approval No
Autoclavable Yes, max. Temperature 130 °C CIP Yes Steam Sterilizable Yes, max. Temperature 140 °C Operating Temperature Range Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C Pressure Range bar g 0 to 20 bar (130 °C)



Marca: Hamilton
+54 9 11 6713-8305